The Laureate™
Weight Meter is more compact and less expensive than
available alternatives. It provides a choice of two signal
conditioner boards: the Laureate load cell signal
conditioner board, which provides full-scale ranges of 20,
50, 100, 250 and 500 mV with 4- or 6-wire load cell hookup,
and the Laureate DC signal conditioner board, which allows
the meter to accept preconditioned signals such as 0-10 V or
4-20 mA. Options include setpoint control, analog output and
digital communication boards.
PLEASE SEE TECHNICAL INFORMATION - PDF FOR COMPLETE
INFORMATION |
|
Features |
|
|
Specifications |
|
With
Load Cell Signal Conditioner |
|
Full-Scale
Input, mV |
Zero
Adjust |
Span
Adjust |
Error at 25°C |
|
±20.000
±50.000
±100.00
±250.00
±500.00 |
-99,999
to
+99,999 |
0 to
±99,999 |
0.0030% of
reading /°C |
|
|
With
DC Signal Conditioner |
|
Full-Scale
Input, mV |
Zero
Adjust |
Span
Adjust |
Error at 25°C |
|
±200 mV
±2 V
±20 V
±20 mA |
-99,999
to
+99,999 |
0 to
±99,999 |
0.01% FS
±1 count |
|
|
With
Load Cell or DC Signal Conditioners |
|
Accuracy, Load Cell Signal |
Span Tempco
Zero Tempco
Calibration Method |
0.0030% of reading/°C
Auto-zero
2 points using zero weight and known weight |
|
Accuracy, DC Signal |
Span Tempco
Zero Tempco
Calibration Method |
0.01% FS ±1 count
Auto-zero
2 points using zero weight and known weight |
|
Noise Rejection |
CMR, DC to 60 Hz
NMR at 50/60 Hz |
130 dB
90 dB with min filtering |
|
A-to-D Conversion |
Technique
A-to-D Rate
Output Update
Display Update |
Concurrent Slope™
(Pat 5,262,780)
60/s at 60 Hz, 50/s at 50 Hz
56/s at 60 Hz, 47/s at 50 Hz
3.5/s at 60 Hz, 3/s at 50 Hz |
|
Display |
Readout
Color
Range
Rounding
Indicators |
5 digits, 7-segment,
14.2 mm (.56")
Red or green LED
-99999 to +99999
-999990 to +999990 (with fixed selectable
right-hand zero)
Count by 1, 2, 5, 10, 20, 50, 100
Four LED lamps |
|
Power |
Voltage, std.
Voltage, opt.
Frequency
Power isolation |
85-264 Vac and 90-370 Vdc
8-28 Vac and 9-37 Vdc
DC or 49-440 Hz
Safety-rated to 250 Vac, meter ground to
earth ground,
DC to 60 Hz, 4.2 kVp per High Voltage Test |
|
Excitation Output |
5 Vdc
10 Vdc
24 Vdc
Output isolation |
5 Vdc ±5%, 100 mA max
10 Vdc ±5%, 120 mA max
24 Vdc ±5%, 50 mA max
50 Vdc to meter ground |
|
Environmental |
Operating Temperature
Storage Temperature
Relative Humidity
Protection |
0°C to 55°C
-40°C to 85°C
95% at 40°C, non-condensing
NEMA-4X (IP-65) when panel mounted |
|
|
|
|
|
|
. |
|